Shubha, V and Ramesh, TG (1995) Fabrication of Thermo power Analyser for High Temperature Studies. Technical Report. National Aerospace Laboratories, Bangalore, india.
![]() |
PDF
tr_pdmt_9508_R80770_txt.pdf Restricted to Repository staff only Download (1MB) |
![]() |
Indexer Terms (Generate index codes conversion from application/pdf to indexcodes)
indexcodes.txt Restricted to Repository staff only Download (2kB) |
Abstract
A totally automated absolute thermoelectric power (TEP) measuring :system for studies at high temperatures (up to 800xB0;C) has been fabri- cated based on the design developed by the authors earlier. The main feature of this system is the control of the mean temperature of the sample to within xB1;_0.1 xB0; C over the entire tempe- rature range. The resolution in absolute TEP measurement is =0.1 Vv/xB0;C. Some typical data on the behaviour of TEP across the magnetic13; phase transition in nickel is presented. This system has been installed at the Inter University Consortium, Indore.
Item Type: | Monograph (Technical Report) |
---|---|
Uncontrolled Keywords: | Thermopower Analyser |
Subjects: | CHEMISTRY AND MATERIALS > Composite Materials CHEMISTRY AND MATERIALS > Chemistry and Materials (General) |
Depositing User: | Mrs Neetu Chandra |
Date Deposited: | 31 May 2006 |
Last Modified: | 24 May 2010 04:12 |
URI: | http://nal-ir.nal.res.in/id/eprint/1586 |