Temperature programmer for the teflon-cell arrangement in high pressure studies

Shubha, V and Ramesh, TG (1986) Temperature programmer for the teflon-cell arrangement in high pressure studies. Technical Report. National Aeronautical Laboratory, Bangalore, India.

[img] PDF
tr_pd_tm_mt_8602_txt.pdf
Restricted to Repository staff only

Download (4MB)

Abstract

A precision analog temperature programmer is presented for use with the teflon cell technique which is extensively .used in high pressure studies. The stability of temperature control is within 10.1xB0;C of the13; set point value. Linear temperature versus time ramps selectable from 1xB0;C/min to 20xB0;C/min is also possible in this system. This instrument facilitates automatic data acquisition of electronic transport properties under high pressure conditions.